Vous avez été redirigé vers la version locale de la page demandée

AN-I-036

2025-04

Sodium content in water using an ion-selective electrode

Fast, accurate, and selective determination in mineral water and leachate according to AOAC 976.25


Summary

Groundwater naturally occurs by means of precipitation, and it gathers dissolved minerals as it filters and passes through the soil. Typically, groundwater is used as a source for drinking water and irrigation purposes.

Rain and snow can also leach various substances, such as sodium salts, from landfills into groundwater. This so-called leachate is considered hazardous to the environment and can contaminate groundwater reserves.

This Application Note describes sodium analysis in water (mineral water as well as leachate) using the separate sodium-selective electrode, also known as the Na-ISE. All samples in this study were determined by standard addition. The method is based on the standard AOAC 976.25.


Configuration


Introduction

The standard addition (STDADD) is recommended for undefined or complex sample matrices. In the standard addition method, a defined amount of the ion of interest is added to a known volume of sample (in several steps). The unknown concentration can be calculated from the resulting potential differences between the sample and the sample with added standard solution. This calculation is performed automatically by modern ion meters or software such as OMNIS.


Sample and Sample Preparation

The analysis is demonstrated on mineral water and leachate.

No sample preparation is required.


Experimental

OMNIS Advanced Titrator and OMNIS Dosing Module
Figure 1. OMNIS Advanced Titrator and OMNIS Dosing Module equipped with Na-ISE for the determination of sodium.

An OMNIS Advanced Titrator and an OMNIS Dosing Module equipped with a separate sodium-selective electrode were used to measure sodium in water samples (Figure 1).

To a reasonable amount of sample, ISA solution consisting of c(CaCl2) = 1 mol/L is automatically added and the standard addition is carried out with the sodium standard solution β(Na+) = 200 mg/L or β(Na+) = 2000 mg/L, respectively. The mineral water is measured without ISA addition, as this is not necessary if the concentration in the sample solution is <50 mg/L Na+.


Results

Standard addition provides accurate and reproducible results with SD(rel) < 2.0% as shown in Table 1. An example of standard addition is shown in Figure 2.

Table 1. Results for sodium in mineral water and leachate by standard addition.
Sample (n = 6) Mean value SD(abs) SD(rel) in %
Mineral water 5.08 mg/L  0.00 mg/L 0.1
Leachate 151.1 mg/kg 0.4 mg/kg 0.3
Results for sodium in mineral water and leachate
Figure 2. Standard addition of mineral water with four increments made with 200 mg/L sodium.

Conclusion

Results of sodium measurement in water with the sodium ion-selective electrode are fast, reproducible, and accurate. In principle, this method can be used for any type of wastewater or sewage water for different sodium concentrations. Appropriate dilution or clarification (e.g., filtration or centrifugation) may be required prior to analysis.

Handling the Na-ISE is user friendly and there is no need to condition the electrode, which means it is immediately ready for use. Furthermore, this method conforms to AOAC 976.25.

The analytical system presented in this application offers users flexibility combined with high-end software. The measuring range of the separate polymer Na-ISE lies between 5 × 10–6 mol/L and 1 mol/L Na+ (corresponds to approximately 0.11 mg/L Na+) and is suitable for a wide range of different samples, from foodstuffs to pharmaceutical products and cosmetics.

As well as improving the precision and speed of sodium determinations, OMNIS delivers results that are equal to or better than other established titration systems.

Contact

Metrohm Suisse SA

Industriestrasse 13
4800 Zofingen

Contact