You have been redirected to your local version of the requested page

This Application Note describes the determination of N,N-diethylhydroxylamine (DEHA), triisopropanolamine (TIPA) and a cationic color developing component (CDC) in a developer solution. The analysis is performed on a high-capacity column such as Metrosep C - 250/4.0 with subsequent direct conductivity detection. To reduce the residence time of the strongly retained color developer components, the column flow rate is increased after the elution of the amines.

Contact

Metrohm Vietnam Co., LTD

Park IX Building 08 Phan Dinh Giot Street, Ward Tan Son Hoa, Tan Binh District
700000 Ho Chi Minh City

Contact