AN-S-298

通过样品重注技术测定高氯和高硫酸盐基质中的砷酸盐


概要

The anion column Metrosep A Supp 15 - 150/4.0 is a pretty high-capacity column. The direct determination of arsenate in a matrix of 180 mg/L chloride and 320 mg/L sulfate is not possible, as the arsenate is hardly detectable under the sulfate peak tail. Sample reinjection cuts off the majority of the matrix and therefore allows an accurate determination of the arsenate.