Nickel and cobalt in zinc plant electrolytes (concentrated zinc sulfate solutions)


Tóm tắt

Nickel can be determined in concentrated zinc solutions by adsorptive stripping voltammetry (AdSV) at the HMDE using ammonia buffer as supporting electrolyte and dimethylglyoxime (DMG) as complexing agent. The determination of cobalt does not work under these conditions as the very high Zn2+ concentration interferes with the Co signal. Therefore, an alternative complexing agent has to be used: α-benzil dioxime in ammonia buffer under addition of sodium nitrite.