AN-V-158

Indium in a tin bath


Summary

The concentration of In in a Sn bath is determined in a HCl / Urotropin® containing electrolyte by anodic stripping voltammetry (ASV). The determination is linear up to approx. 0.5 mg/L with respect to the concentration of In in the measuring vessel. The standard addition solution is also prepared with HCl and Urotropin®.

Kontakt

Metrohm Nordic AB

Box 11065
161 11 Bromma

Kontakt

Find Applications
Förfina din sökning

Applikationen har taggats under

// Plating baths – other metals