Zostałaś(eś) przekierowany do lokalnej wersji strony

AN-V-111

Germanium in zinc plant electrolytes (concentrated ZnSO4 solutions)


Summary

Germanium is determined by adsorptive stripping voltammetry (AdSV) at the HMDE using aqueous sulfuric acid as supporting electrolyte and pyrocatechol violet as complexing agent. It is possible to determine 20 µg/L Ge in a sample containing 150 g/L Zn, 3 g/L Cd and 1 mg/L Pb.

Kontakt

Metrohm Polska

ul. Centralna 27
05-816 Opacz-Kolonia