AN-T-149

Determination of thorium using automatic photometric titration


Summary

Thorium is titrated with EDTA at a pH value of 4.9. Xylenol orange is used as the indicator for visualization of the equivalence point. The equivalence point is determined with the Optrode at a wavelength of 574 nm.