요청하신 페이지의 적용되는 국가로 리디렉션되었습니다

AN-V-109

Selenium in wastewater after UV digestion


Summary

Selenium is determined by cathodic stripping voltammetry (CSV) at the hanging mercury drop electrode (HMDE). Se(IV) is deposited on the surface of the mercury drop in sulfuric acid electrolyte under addition of copper ions as Cu xSe y.

Wastewater samples containing organic contaminants have to be digested by UV irradiation before analysis. In addition, the sample has to undergo a second irradiation step at pH 7−9 to reduce Se(VI) to Se(IV), since only Se(IV) is electrochemically active.

 

 

 

문의하기

Metrohm AG

Ionenstrasse
9100 Herisau

문의하기

애플리케이션 찾기
선택 후 재검색

이 애플리케이션의 태그는 다음과 같습니다

// Water – wastewater