You have been redirected to your local version of the requested page

AN-C-156

Temperature dependency of the cation separation on the Metrosep C 6 - 150/4.0 column


Summary

The column temperature influences the duration of the cation separation on the high-performance Metrosep C 6 - 150/4.0 column. The retention times of lithium, sodium, ammonium, magnesium, and calcium remain practically constant with increasing column temperature, whereas those of potassium, strontium, and barium are considerably shortened. This means that the temperature can be used to reduce analysis time considerably on the Metrosep C 6 - 150/4.0.

Contact

Metrohm SG

31 Toh Guan Road East, LW Technocentre, #06-08
608608 Singapore

Contact

Find Applications
Refine your search

This application has been tagged under

// Education & basic research