AN-T-148

Determination of zirconium using automatic photometric titration


Summary

Zirconium is titrated directly with EDTA in acidic aqueous solution (buffer, pH 1). Eriochrome cyanine R is used as the indicator for this procedure. The equivalence point is determined with the Optrode at a wavelength of 520 nm.