Application Finder
- AN-COR-010Electrochemical Corrosion Studies of Various Metals
Corrosion of metals is a problem seriously affecting not only many industrial sectors, but also private life, resulting in enormous costs. In this application note, the results gained during electrochemical corrosion studies on different metals are compared to literature data.
- AN-V-193Mercury in metallic materials as part of electrotechnical products
The EU directive on «Restriction of Hazardous Substances» (RoHS) requires the testing of four regulated heavy metals (Pb, Hg, Cd, Cr(VI)) in electrotechnical products. After sample preparation according to IEC 62321 the determination of mercury in metallic materials can be carried out by anodic stripping voltammetry (ASV) at a gold rotating disk electrode (Au-RDE).
- AN-V-191Cadmium and lead in metallic materials as part of electrotechnical products
The EU directive on «Restriction of Hazardous Substances» (RoHS) requires the testing of four regulated heavy metals (Pb, Hg, Cd, Cr(VI)) in electrotechnical products. After sample preparation according to IEC 62321 the determination of lead and cadmium in metallic materials can be carried out by anodic stripping voltammetry (ASV) using ammonium oxalate buffer pH 2.
- AN-V-192Chromium(VI) in chromate coating on metallic materials as part of electrotechnical products
The EU directive on «Restriction of Hazardous Substances» (RoHS) requires the testing of four regulated heavy metals (Pb, Hg, Cd, Cr(VI)) in electrotechnical products. After sample preparation according to IEC 62321 the determination of chromium(VI) in chromate coating on metallic materials can be carried out by adsorptive stripping voltammetry (AdSV) using DTPA (diethylenetriamine pentaacetic acid) as complexing agent.
- AB-147Simultaneous trace determination of seven metals in «electronic grade» materials using stripping voltammetry
The metals Cd, Co, Cu, Fe, Ni, Pb, and Zn are determined in the sub-ppb range (limit of detection 0.05 µg/L) by means of stripping voltammetry. The DP-ASV method is used for Cd, Cu, Pb, and Zn whereas Co, Ni, and Fe are determined by means of the DP-CSV method (dimethylglyoxime or catechol complexes).Use of the VA Processor and the sample changer allows automatic determination of the above metal ions in one solution. The method has been specially developed for trace analysis in the manufacture of semiconductor chips based on silicon. It can naturally also be employed successfully in environmental analysis.
- AN-U-059Transition metals in feedwater using post-column reaction and subsequent UV/VIS detection
Feedwater for steam generation in boiling water reactors (BWR) needs to be analyzed for corrosion products. Presence of transition metals, mainly nickel and iron, indicates corrosion problems. Traces of these ions are determined using Inline Preconcentration (MiPCT). After separation, post-column reaction with 4-(2-pyridylazo)resorcinol (PAR) allows VIS detection at 510 nm.
- AN-S-108Five anions in an electroplating bath after inline elimination of heavy metals
Determination of chloride, nitrite, nitrate, phosphate, and sulfate in an electroplating bath after inline elimination of heavy metals by cation exchange on the 793 IC Sample Prep Module using anion chromatography with conductivity detection after chemical suppression.
- AN-U-040Chromium(VI) in colorless and colored chromate coating on metallic samples as per IEC 62321 method for RoHS testing
The determination of chromium in metal plate samples using anion exchange chromatography with UV/VIS detection after post-column reaction with diphenylcarbazide as per IEC 62321 method for RoHS testing. This method provides procedures for the determination of the presence of chromium(VI) in colorless and colored chromate coatings on metallic samples.
- AN-U-067Trace determination of transition metals with preconcentration (MiPCT), post-column reaction, and UV/VIS detection
The determination of transition metals by ion chromatography is possible with direct conductivity detection (see AN-C-137) as well as with UV/VIS detection after post-column reaction. Here, the cations are separated as anionic complexes and analyzed after post-column reaction with PAR with subsequent UV/VIS detection. Speciation determination of iron (separation of Fe(II) and Fe(III)) is possible with this procedure. For trace analysis, Metrohm Inline Preconcentration Technique (MiPCT) is applied.
- 8.000.6086Semi-continuous determination of anions, cations, and heavy metals in aerosols using PILS-IC-VA
This poster presents an approach that couples a Particle-Into-Liquid-Sampler (PILS) to a dual-channel ion chromatograph (IC) for measurement of aerosol anions and cations and a voltammetric measuring stand (VA) to determine the heavy metals. Feasibility of the PILS-IC-VA online system was demonstrated by collecting aerosol samples in Herisau Switzerland, at defined time intervals; air pollution events were simulated by burning lead- and cadmium-coated sparklers.
- AN-COR-018Evaluation of organic coatings on metals using Autolab PGSTAT based on ISO 17463 – Paints and Varnishes
The International Standard ISO 17463 describes the determination of the anticorrosive properties of high impedance organic protective coatings on metals. This technique uses cycles composed of electrochemical impedance spectroscopy (EIS) measurements, cathodic polarizations and potential relaxation. This application note shows the compliance of the Metrohm Autolab PGSTAT M204 and flat cell with the standard ISO 17463.
- AB-112Quantitative determination of metals that can be precipitated by potassium hexacyanoferrate(II) in wine («décassage» of wine)
Wine sometimes contains heavy metals which can be precipitated out by the addition of potassium ferrocyanide. Generally, these are quantities of iron ranging between 1 and 5 mg, and exceptionally up to 9 mg Fe/L. Zinc, copper, and lead – in descending order of content – may also be present. To estimate the quantity of potassium ferrocyanide necessary for the «décassage of the wine», only very complicated and relatively inaccurate methods have been described until now.This Bulletin permits accurate results to be obtained easily with a simple instrumentation. The results are available in a short time.
- AN-U-052Transition metal cations applying UV/VIS detection after post-column reaction with PAR (4-(2-pyridylazo)resorcinol)
The Metrosep C 4 columns are mainly used for the separation of alkali and alkaline earth metal cations including ammonium and organic amines. Additionally transition metals may be determined.
- AB-036Half wave potentials of metal ions for the determination by polarography
In the following tables, the half-wave potentials or peak potentials of 90 metal ions are listed. The half-wave potentials (listed in volts) are measured at the dropping mercury electrode (DME) at 25 °C unless indicated otherwise.
- AN-EC-015Metrohm 663 VA stand for Heavy Metal Ions detection in Water Sample
The determination of heavy metal ions in a solution is one of the most successful application of electrochemistry. In this application note, anodic stripping voltammetry is used to measure the presence of two analytes, in a sample of tap water.
- AN-H-142Determination of metal-organic compounds
Metal-organic compounds are commonly used in organic chemistry, for example as Grignard reagents or as strong bases (e.g., butyl lithium compounds). The knowledge of the exact content of reactive species allows to better plan the required amounts for reactions preventing the waste of material or too low yields.This Application Note describes the analysis of metal organics by thermometric titration using 2-butanol as titrant. Due to the strongly exothermic nature of the reaction between 2-butanol with metal-organic compounds, a fast and quantitative analysis of these substances is possible.
- AN-T-024Metal contents of alkaline plating baths for cadmium, copper, lead or zinc
Determination of cadmium, copper, lead, and zinc in alkaline plating baths by potentiometric titration with EDTA using the Cu-ISE.
- AN-H-105Determination of nickel in hydrometallurgical leach liquors
A measured amount of acidic hydrometallurgical leach liquor is first treated with hydrogen peroxide to oxidize Fe(II) to Fe(III), then with potassium pyrophosphate solution to mask interference from Fe(III) and other metal ions. Ammonium acetate solution is then added as a pH modifier, before being titrated with standardized disodium dimethylglyoximate to an exothermic endpoint.
- AN-V-104Formaldehyde in metalworking lubricants
Formaldehyde is determined polarographically at the DME in alkaline solution.
- AB-101Complexometric titrations with the Cu ISE
This Bulletin describes the complexometric potentiometric titration of metal ions. An ion-selective copper electrode is used to indicate the endpoint of the titration. Since this electrode does not respond directly to complexing agents, the corresponding Cu complex is added to the solution. With the described electrode, it is possible to determine water hardness and to analyze metal concentrations in electroplating baths, metal salts, minerals, and ores. The following metal ions have been determined: Al3+, Ba2+, Bi3+, Ca2+, Co2+, Fe3+, Mg2+, Ni2+, Pb2+, Sr2+, and Zn2+.
- 8.000.6059Determination of hazardous substances in electrical and electronic equipment
The Restriction of Hazardous Substances (RoHS) Directive 2002/95/EC stipulates maximum limits for the hazardous metals cadmium, lead and mercury as well as the hexavalent chromium and the brominated flame retardants in electrical and electronic products. To ensure compliance, reliable analysis methods are required.This poster deals with the wet-chemical determination of trace concentrations of the six RoHS-restricted substances in a wide variety of materials including metals, electrotechnical components, plastics and wires. After sample preparation according to IEC 62321, the metals lead, cadmium and mercury are best determined by anodic stripping voltammetry (ASV) and the flame retardants PBB and PBDE are quantified by direct-injection ion chromatography (IC) using spectrophotometric detection. Chromium(VI) can be determined either by adsorptive stripping voltammetry (AdSV) or IC. Both methods are very sensitive and meet prescribed RoHS limits.
- AN-H-044Standardization of EDTA titrant by copper
Thermometric complexometric titration of metals is often performed with tetrasodium EDTA. This Application Note explains the standardization of tetrasodium EDTA titrant with copper.
- AN-H-046Standardization of copper back-titrant by EDTA
Standardization of copper back-titrant using standard tetrasodium EDTA titrant in the determination of metals.
- 8.000.6015Effect of eluent composition and column temperature on IC column retention times
This work was carried out with a Metrosep C 2 - 150 separation column, the following eluent parameters being investigated: nitric, tartaric, citric and oxalic acid concentration and concentration of the complexing anion of dipicolinic acid (DPA). The aim was to determine the effect of these parameters plus that of the column temperature on the retention times of alkali metals, alkaline earth metals, ammonium and amines using ion exchange chromatography with non-suppressed conductivity detection. Due to similar affinities for the ion exchange column, transition metals are difficult to separate with the classical nitric, tartaric, citric and oxalic acid eluents. Partial complexation with the dipicolinate ligand significantly shortens the retention times and improves the separation efficiency. However, too strong complexation results in a rapid passage through the column and thus in a complete loss of separation. Apart from a change in the elution order of magnesium and calcium at high DPA concentrations, other non-amine cations are only slightly affected by the eluent composition. Irrespective of the tartaric acid and nitric acid concentration in the eluent, an increase in column temperature shortens the retention times and slightly improves the peak symmetries of organic amine cations, particularly in the case of the trimethylamine cation. In contrast, an increase in column temperature in the presence of DPA concentrations exceeding 0.02 mmol/L increases the retention time of the transition metals. Depending on the separation problem, variation of the pH value, the use of a complexing agent and/or an increase in column temperature are powerful tools for broadening the scope of cation chromatography.
- AN-H-107Determination of ferrous ion content in hydrometallurgical leach liquors
A measured amount of acidic hydrometallurgical leach liquor is further acidified with sulfuric acid, prior to being titrated with standard potassium dichromate solution to an exothermic endpoint. Thus, 1 mol K2Cr207 ≡ 6 mol Fe2+.
- AN-H-108Determination of ferric ion in hydrometallurgical leach liquors
A measured amount of acidic hydrometallurgical leach liquor is pH modified with a small amount of glacial acetic acid, and the Fe(III) content reduced to Fe(II) with iodide ion. The liberated iodine is titrated with standard thiosulfate solution to an exothermic endpoint. Thus, 1 mol Fe3+= 1 mol S2O32-.
- AN-H-106Determination of magnesium content in hydrometallurgical leach liquors
A measured amount of acidic hydrometallurgical leach liquor is treated first with a complexing agent (sodium gluconate). It is then basified to ~pH 10.5 with a NH3 /NH4Cl buffer, prior to the addition of KCN solution to mask Fe(III). Caution! Do not add KCN to solutions of pH below 9! The Fe(III) is then reduced to Fe(II) by additon of ascorbic acid, prior to titrating the Mg content with standard Na4EDTA solution.
- AN-PAN-1006Online analysis of zinc, sulfuric acid, and iron during zinc refining
This Process Application Note is dedicated to the online analysis of zinc, iron and sulfuric acid in several stages of the zinc production process. Additionally, traces of germanium, antimony, as well as several transition metals (e.g., Ni, Co, Cu, Cd) can be precisely determined (<50 µg/L) in the purification filtrates and reactor trains.
- 8.000.6040Improved cation separation thanks to a new column material
Comparative measurements show that the new Metrosep C 4 cation column has even better separation characteristics than the previous Metrosep C 2 and Metrosep Cation 1-2 column types. The Metrosep C 4 column has a clearly improved peak shape which leads to a better separation of the individual peaks. Using Metrosep C 4 the number of theoretical plates per meter was noticeably higher than that obtained on the Metrosep C 2 or C 1-2 column. Additionally for standard cations transition metals and amines, the Metrosep C 4 column shows better results with respect to peak shape, peak height, resolution and asymmetry factor. The clearly improved resolution of the C 4 column with its narrow and high peaks achieves baseline separation for six standard and six transition metal cations. Analysis times and peak areas obtained with the C 4 column are in the same range as those obtained with its predecessors.As a result of the latest production methods and materials, the promising Metrosep C 4 column excels by an outstanding separation performance for complex mixtures comprising standard cations, transition metal cations and amines.
- AB-429Determination of copper in water with the scTRACE Gold
Copper is one of the few metals which is available in nature also in its metallic form. This and the fact that it is rather easy to smelt led to intense use of this metal already in the so-called Copper and Bronze Age. Nowadays, it is more important than ever, because of its good electrical conductivity and its other physical properties. For plants and animals, it is an essential trace element; for bacteria, in contrast, it is highly toxic.This Application Bulletin describes the determination of copper by anodic stripping voltammetry (ASV) using the scTRACE Gold electrode. With a deposition time of 30 s, the limit of detection is about 0.5 μg/L.
- AN-PAN-1066Online analysis of cadmium in incineration plant wet gas scrubbers
Incineration flue gas requires treatment such as wet scrubbing. The 2060 VA Process Analyzer monitors heavy metals in the scrubbing water, ensuring compliance.
- WP-087Green alternative methods for voltammetric analysis in different water matrices
This White Paper presents four different «green» sensors: the scTRACE Gold, screen-printed electrodes, the glassy carbon electrode, and the Bi drop electrode from Metrohm that can be used to determine low concentrations of heavy metals in different sample matrices, such as boiler feed water, drinking water, and sea water.
- AB-114Determination of copper, nickel, cobalt, zinc, and iron in a single operation by polarography
Cu2+, Co2+, Ni2+, Zn2+, and Fe2+/Fe3+ are determined simultaneously. Interference due to the presence of other metals is mentioned, and methods given to eliminate it. The threshold of determination is ρ = 20 µg/L for Co and Ni, and ρ = 50 µg/L each for Cu, Zn, and Fe.
- AN-H-047Determination of nickel by EDTA back-titration
Determination of nickel in refinery and plating solutions. When other metals capable of being complexed by EDTA are present, these will interfere and enhance the result for nickel.
- AN-S-301Comprehensive water analysis with VoltIC pro I
VoltIC pro I is the perfect combination of voltammetry and ion chromatography for the fully automated analysis of anions, cations, and heavy metals (e.g., Zn, Cd, Pb, Cu): comprehensive water analysis on a single system.
- AN-T-218Analysis of Li-ion battery cathode materials made from Co, Ni, and Mn
The lithium-ion battery market is continuously growing due to the tremendous demand for battery powered consumer products. So-called «NCMs», a mixture of nickel, cobalt, and manganese oxides, have been gathering interest as cathode materials, replacing traditional compounds like cobalt oxides.Quality analysis of the post-sintered materials or recycled batteries can be performed by titration, as demonstrated in this Application Note. A fully automated analysis of the corresponding metals can be performed with OMNIS and its pipetting equipment.
- AN-S-347Comprehensive water analysis with VoltIC Professional 1
VoltIC Professional 1 is the perfect combination of voltammetry and ion chromatography for the fully automated, simultaneous analysis of anions, cations, and heavy metals (e.g., Zn, Cd, Pb, Cu). The multiple-parameter analysis uses the same "Liquid Handling" elements and a shared sample changer, thus saving on space and costs.
- AN-T-091Fully automatic determination of total content of Ba, Ca, Mg, Pb and Zn in unused lubricating oils
This Application Note describes the determination of total content of Ba, Ca, Mg, Pb and Zn in unused lubricating oil by means of the Optrode (610 nm). An excess of EDTA is first added to the metals. Afterwards, the excess EDTA is titrated back with magnesium chloride solution up to the end point of the indicator Eriochrome Black T.
- AN-V-094Platinum and rhodium in red wine after UV digestion
For the determination of heavy metals in wine, UV digestion is required to mineralize the sample. The determination of platinum and rhodium is carried out with adsorptive stripping voltammetry (AdSV) at the HMDE.
- 8.000.6011Ion chromatographic determination of anions, cations and organic acids in biofuels
Quality and process control of biofuels require straightforward, fast and accurate analysis methods. Ion chromatography (IC) is at the leading edge of this effort. Traces of anions in a gasoline/ethanol blend can accurately be determined in the sub-ppb range after Metrohm Inline Matrix Elimination using anion chromatography with conductivity detection after sequential suppression. While the analyte anions are retained on the preconcentration column, the interfering organic gasoline/bioethanol matrix is washed away.Detrimental alkali metals and water-extractable alkaline earth metals in biodiesel are determined in the sub-ppm range using cation chromatography with direct conductivity detection applying automated extraction with nitric acid and subsequent Metrohm Inline Dialysis. Unlike high-molecular substances, ions in the high-ionic strength matrix diffuse through a membrane into the low-ionic water acceptor solution. In biogas reactor samples, low-molecular-weight organic acids stem from the biodegradation of organic matter. Their profile allows important conclusions concerning conversion in the anaerobic digestion reaction. Volatile fatty acids and lactate can be accurately determined by using ion-exclusion chromatography with suppressed conductivity detection after inline dialysis or filtration.
- AN-CS-007Lithium in addition to other cations in seepage water from minerals through sequential suppression
With the advent of electric automobiles, the demand for lithium batteries and with it the demand for lithium material will increase sharply. Brine lakes and hard silicate minerals are numbered among the most important sources of lithium. This Application Note addresses cation determination in seepage water from lithium minerals. Alkali and earth alkali metals are separated in the lithium digestions on the Metrosep C Supp 1 - 250/4.0 column, with subsequent conductivity detection after sequential suppression.
- AN-V-186Chromium(VI) in electronic components as part of electrotechnical products
The EU directive on «Restriction of Hazardous Substances» (RoHS) requires the testing of four regulated heavy metals (Pb, Hg, Cd, Cr(VI)) in electrotechnical products. After sample preparation according to IEC 62321 the determination of chromium(VI) in electronic components can be carried out by polarography in ammonia buffer pH 9.6.
- AN-V-189Chromium(VI) in polymer materials as part of electrotechnical products
The EU directive on «Restriction of Hazardous Substances» (RoHS) requires the testing of four regulated heavy metals (Pb, Hg, Cd, Cr(VI)) in electrotechnical products. After sample preparation according to IEC 62321 the determination of chromium(VI) in polymer materials can be carried out by polarography in ammonia buffer pH 9.6.
- AN-V-223Nickel and cobalt in drinking water with a Bi drop electrode
The main sources of nickel pollution are electroplating, metallurgical operations, or leaching from pipes and fittings. Catalysts for the petroleum and chemical industries are major application fields for cobalt. In both cases, the metal is either released directly, or via the waste water-river pathway into the drinking water system. Therefore in the EU the legislation specifies 20 µg/L as the limit value for the Ni concentration in drinking water.The simultaneous and straightforward determination of nickel and cobalt is based on adsorptive stripping voltammetry (AdSV). The unique properties of the non-toxic Bi drop electrode combined with AdSV results in an excellent performance in terms of sensitivity.
- AN-C-157Sodium and potassium in polyol using IC following inline matrix elimination
Polyols are important raw materials in polyurethane production. Contamination in the raw materials have a great influence on reactions and impair the quality of the end product. Alkali metals are particularly strong catalysts for linear or branched reactions. A rapid and precise method for their simultaneous determination is ion chromatography following Inline Matrix Elimination.
- AN-H-118Determination of hydrochloric acid in acidic solutions containing iron and aluminum
The presence of the hydrated ion [Fe(H2O)6]3+ can interfere with the determination of «free acid» due to the low pKa value (~2.2) of this ion. Ions of metals such as Fe, Cu, and Al can be masked effectively with fluoride, and permit the determination of the acid content by thermometric alkalimetric titration with good accuracy and precision.
- AN-V-185Cadmium and lead in electronic components as part of electrotechnical products
The EU directive on «Restriction of Hazardous Substances» (RoHS) requires the testing of four regulated heavy metals (Pb, Hg, Cd, Cr(VI)) in electrotechnical products. After sample preparation according to IEC 62321 the determination of lead and cadmium in electronic components can be carried out by anodic stripping voltammetry (ASV) using ammonium oxalate buffer pH 2.
- WP-021Water analysis in the field: Determining arsenic, mercury, and copper
Heavy metals such as arsenic and mercury find their way into the ground water in many regions of the world, either through natural processes or as the result of human activities. Limit values are exceeded many times over, particularly for arsenic in drinking water, in many areas. This calls for a rigorous monitoring of water quality. The present whitepaper focuses on field determinations of arsenic, mercury, and copper – directly at the sampling site.
- AN-V-187Mercury in electronic components as part of electrotechnical products
The EU directive on «Restriction of Hazardous Substances» (RoHS) requires the testing of four regulated heavy metals (Pb, Hg, Cd, Cr(VI)) in electrotechnical products. After sample preparation according to IEC 62321 the determination of mercury in electronic components can be carried out by anodic stripping voltammetry (ASV) at a gold rotating disk electrode (Au-RDE).
- AN-V-188Cadmium and lead in polymer materials as part of electrotechnical products
The EU directive on «Restriction of Hazardous Substances» (RoHS) requires the testing of four regulated heavy metals (Pb, Hg, Cd, Cr(VI)) in electrotechnical products. After sample preparation according to IEC 62321 the determination of lead and cadmium in polymer materials can be carried out by anodic stripping voltammetry (ASV) using ammonium oxalate buffer pH 2.